THE SERIES LCI132 PROCESS INDICATORS offer flexibility and value in a low cost, compact 1/32 DIN package. The shallow depth of these full size panel meters allows installation in panels only 2.76″ (70 mm) deep with room to spare.
Product Applications
- Display process values from various transmitters
LCI132-0X
Range : Input Volts (DC): ±200 V, ±20 V, ±10 V; Input Amperes: ±100 mV, ±20 mA;
Resolution : Input Volts (DC): 0.1 V, 0.01 V, 1 mV; Input Amperes: 0.1 mV, 0.01 mA; Input Impedance: Volts: 1 M Ω, mV: 100 M Ω, mA: 12.1 Ω.
LCI132-1X
Range AC : Input Volts: 600.0, 100.0; Input Amperes: 5.000, 1.000;
Range DC:Input Volts: -199.9, ±600.0, ±100.0; Input Amperes: -1.999, +5.000, ±1.000;
Resolution:Input Volts: 0.1 V; Input Amperes: 1 mA Input Impedance: Volts: 3 M Ω (106), Amps: 12 m Ω (10-3).
Accuracy at 23ºC ±5ºC:100/600 VDC 1/5 A DC; 600V/5A AC: ±(0.2% reading + 3 digits). 100 V / 1A AC: ±(0.4% reading + 4 digits).
Temperature Coefficient : 100 ppm/°C.
Warm Up : 5 minutes.
Power Supply and Fuses (DIN 41661) (Not supplied)
LCI132-X0:85–265 VAC 50/60 Hz and 100-300 VDC: Fuse: 0.1A/ 250V.
LCI132-X1:21-53 VAC 50/60Hz and 10.5-70 VDC: Fuse 0.5A/ 250V.
Display
LCI132-X0:85–265 VAC 50/60 Hz and 100-300 VDC: Fuse: 0.1A/ 250V.
LCI132-X1:21-53 VAC 50/60Hz and 10.5-70 VDC: Fuse 0.5A/ 250V.
Environmental
Operating Temperature : -10 to 60°C.
Storage Temperature : -25 to 85°C.Relative Humidity (non condensed):< 95% @ 40°C.
Panel Sealing : NEMA 4X (IP66).