The GSM-B5SA-BN00 is a low-flow version of the Vögtlin red-y smart GSM thermal mass flow meter, calibrated on air for a working range of 100–5000 mln/min (full scale 5000 mln/min). It reads gas mass flow directly, so there is no need to correct for line pressure or temperature, and it outputs the value as both an analog signal and Modbus RTU.
How It Works
Both red-y families use a thermal MEMS sensor. A small heated element sits in the gas stream, and the gas flowing past carries heat away in proportion to its mass flow. Because the measurement is of mass rather than volume, readings stay stable as pressure and temperature change, and every instrument is calibrated with real gas.
Key Features
- Thermal MEMS mass flow measurement, insensitive to pressure and temperature changes
- Accuracy ±1.0% FS (Standard) or ±0.3% FS + ±0.5% of reading (Hi-Performance, on suitable ranges)
- Analog output (0/4–20 mA, 0/1–5 V or 0/2–10 V) plus RS-485 Modbus RTU as standard
- Optional Profibus, Profinet RT and EtherCAT interfaces
- Multigas option: up to 10 gases or mixtures in one instrument
- Response time about 80 ms; no inlet straight section required
- Free get red-y software for set-up, monitoring and data logging
- 3-year Vögtlin warranty
Applications
- Laboratory gas supply, sensor and detector test benches, and small burner or pilot-flame control
- OEM machines and analysers that need a digital interface
- Semiconductor, pharma and food gas supply lines
Frequently Asked Questions
What flow range does the GSM-B5SA-BN00 cover?
It is calibrated on air for 100–5000 mln/min. The full scale is 5000 mln/min, and the lower figure is the bottom of the guaranteed range.
Can the GSM-B5SA-BN00 measure gases other than air?
Vögtlin calibrates with real gas for air, O₂, N₂, He, Ar, CO₂, H₂, CH₄ and C₃H₈, with other gases and mixtures on request, and the Multigas option stores up to 10 gases in one instrument. Tell us the gas when you order.
What outputs does it have?
An analog output (0–20 mA, 4–20 mA, 0–5 V, 1–5 V, 0–10 V or 2–10 V) and RS-485 Modbus RTU as standard, with Profibus, Profinet RT or EtherCAT as options.
Do you need a controller instead of a meter?
The red-y smart controller GSC has the same sensor plus a built-in control valve.
Celectric is a Vögtlin distributor in Malaysia, based in Penang. Contact us for pricing, gas calibration options and lead time.